What Is the Between-Patient Handling for Tympanometry Tips in Canadian Clinics?

What Is the Between-Patient Handling for Tympanometry Tips in Canadian Clinics?

Tympanometry is the short test in the middle of a hearing appointment: a probe seals against the ear canal, a tone plays, and the middle ear answers. The part nobody prices is the tip. It touches the canal, comes off, and either goes in the waste or back on the probe, and the two choices carry different cleaning obligations, different consumable costs and, in Canada, genuinely different answers depending on which provincial regulator you read. This page sets out what the tip touches, what the manufacturers actually instruct, where the Canadian guidance conflicts with itself, and what a clinic can document so that the answer holds up in an inspection. What the clinic reorders every week is rarely the probe: it is the box of nitrile examination gloves that goes on before the tip and the sealed tips that come off after it.

What does a probe tip actually touch?

The Canadian answer starts with the Spaulding classification, and Ontario's audiology regulator applied it directly: the College of Audiologists and Speech-Language Pathologists of Ontario names impedance probe tips as semi-critical equipment and requires that semi-critical reusable equipment be pre-cleaned and then sterilized or high-level disinfected as a minimum. The same guidelines define a single-use item as one for a single client, and state that items labelled single-use should not be reprocessed. Read together, those two sentences describe the ordinary practice: a sealed tip used once and discarded is not a reprocessing problem at all.

The complication is that the same college published a pandemic-era table placing immittance tips in the non-critical column, where cleaning plus low-level disinfection is the floor. Two documents from one regulator, two classifications, one device. Provincial infection control guidance gives the tie-break: where the level of reprocessing recommended by the manufacturer does not agree with Spaulding's criteria, the higher level must be used. That rule is what turns a disagreement into a defensible operating position, because it removes the option of choosing the less demanding reading in the absence of a reason.

Schematic placing tympanometry equipment on the Spaulding scale: the probe tip as semi-critical, the probe body and headband as non-critical, and the level of reprocessing each one needs.
The item is classified, not the appointment
The classification is applied to the item, not the appointment. One probe can therefore create two obligations at once: the tip is semi-critical, the body it clips onto is non-critical.
What the equipment touches Spaulding class Minimum reprocessing Audiology examples
Intact skin only, or nothing at all Non-critical Cleaning, then low-level disinfection Headbands, bone oscillators, response buttons, the probe body
Mucous membrane or non-intact skin, no penetration Semi-critical Cleaning, then high-level disinfection; sterilization preferred Impedance probe tips, insert earphone tips, otoscope tips
Penetrates the skin or enters a sterile body area Critical Cleaning, then sterilization Not normally encountered in routine audiometry

Are the tips single-use, and what happens when they are reused?

For the three instrument families a Canadian clinic is most likely to own, the manufacturer's answer is the same. The Grason-Stadler TympStar reference manual states that ear tips are single-use only, and that the disposable tip should be pushed firmly onto the probe until fully seated. Interacoustics carries a no-reuse marking on the AT235 and states that its Sanibel ear tips are single use only and should not be reused, with reuse linked to the spread of infection from patient to patient. Audioscan's real-ear work practice control instructs the clinician to fit new single-use probe tubes for every patient and to clean and disinfect the probe dock between patients, including the clip and cable.

Two consequences follow. First, the tip count is not an inventory curiosity: it is the patient count, so a clinic doing forty tympanograms a week consumes forty tips a week whether or not the box was reordered. Second, "single-use" does not mean the hardware behind it is untouched. The probe body, the clip and the cable contact the clinician's hands and the patient's skin and still need a between-patient wipe, which is why the same manufacturer instructions that say "new tip" also say "disinfect the dock".

A cleaning detail worth separating from the infection question: several manufacturers instruct daily cleaning of the probe tubes with a cleaning floss to keep cerumen out of the measurement path. That is an accuracy instruction, not an infection control one, and a clinic that treats the daily floss as its between-patient measure has answered the wrong question. The floss protects the reading; the new tip protects the next patient.

What has to happen between two patients?

The sequence is short enough to be written on a card, which is the point: a procedure that cannot be posted gets done from memory and then varies by clinician. Hand hygiene first. Fit a new tip from the sealed pack in front of the patient, not from a loose handful in a tray. Run the test. Remove the tip and discard it into the ordinary clinical waste stream rather than back into a pocket or a drawer. Wipe the probe body, clip and cable with the facility's low-level disinfectant and allow the stated contact time — a wipe that is dry in five seconds has not disinfected anything. Then hand hygiene again before the next patient.

Where a clinic has chosen reusable tips instead, the middle of that sequence changes and everything else stays: pre-clean to remove cerumen, then high-level disinfection or sterilization, then dry and store covered. That path needs a documented reprocessing area, a validated disinfectant, and a record of each cycle, which is a materially higher operating burden than a box of disposables. The reason clinics still choose it is usually a supply-chain one — a tip that is out of stock blocks the appointment, whereas a bottle of high-level disinfectant does not.

Schematic of the between-patient sequence for a tympanometry probe: hand hygiene, new single-use tip, test, discard the tip, disinfect the probe body, hand hygiene again.
Six steps, one of which is the whole decision
Five documents, three different answers for one small plastic tip. The order in which a clinic reads them is the whole decision.
Document What it says about the tip Where it leaves room
CASLPO audiology infection control guidelines (2010) Names impedance probe tips as semi-critical and requires cleaning plus high-level disinfection or sterilization as a minimum Does not say whether a disposable tip satisfies the requirement, which is how single-use tips became the default answer
CASLPO General Standards, infection prevention (2020) Anything reusable that contacts mucous membrane, non-intact skin or blood must be disinfected by high-level procedures Silent on single-use items, because they are not reprocessed at all
CASLPO pandemic practice advice (2020) A Spaulding table that places immittance tips in the non-critical column Contradicts the 2010 guideline on the same item — the conflict a clinic is most likely to be caught by
PIDAC clinical office practice, Table 2 Where the manufacturer's level is not in agreement with Spaulding, the higher level must be used Resolves the conflict in one direction only: up, never down
Alberta, College of Hearing Aid Practitioners (2023) Lists tympanometry tips under critical equipment, requiring sterilization after each use or disposal A stricter provincial position than Ontario's for the same device

What do the provinces actually require of a clinic?

The provincial spread is real and a clinic operating in more than one province has to notice it. Ontario's regulator describes the tip as semi-critical and accepts high-level disinfection as the minimum for a reusable one. Alberta's hearing aid practitioner college goes further and lists tympanometry tips under critical equipment, requiring sterilization after each use or disposal after use. A disposable tip satisfies both; a wiped reusable tip satisfies neither if the stricter provincial reading is the one an inspector applies.

The general standard that sits behind both is the same sentence in different words: anything reusable that meets mucous membrane or non-intact skin has to be reprocessed to a high level. That is why the disposable route keeps winning on effort. It converts a reprocessing obligation into a purchasing decision, and the purchasing decision is the easy part — a case of nitrile examination gloves and a box of tips sized to the probe, reordered on the appointment count.

What should be written down, and what belongs on the order?

The record is not a log of every tip. It is a one-page procedure: how many tips per patient, who fits them, what happens to the probe body between patients, which disinfectant and which contact time, and what the clinic does when a tip is unavailable. Add the equipment cleaning schedule that the manufacturer's manual implies, and the file can be answered in a single sentence when someone asks "how do you handle the probe between patients".

On the purchasing side, three lines cover the routine: tips, gloves and a between-patient wipe. The 100-count box of nitrile examination gloves is the unit most single-room audiology practices actually consume, while a procedure mask belongs on the same order wherever the clinic's own policy requires one for a face-to-face appointment. What does not belong on the order is a bulk assumption: tip consumption tracks appointments, and appointment volume in audiology is seasonal in a way that glove consumption is not.

Primary sources

The regulatory position is provincial in Canada, so the documents below are the regulators and the manufacturer instructions rather than a national code.

Where does a practice take this next?

Three of the tools here are usable rather than readable. The printable compliance log carries the fields a receiving check, a lot trace and a reprocessing record actually need, so the paperwork can start without inventing a format. The clinic supply planner turns an appointment count into tips, gloves and wipes on one order, so nothing runs out mid-clinic. The sterilization monitoring calculator works out how many indicators a year of cycles consumes, and the provincial self-check lists what an inspector asks for in each province, and multi-site or purchasing-group orders run through the wholesale account route. Reselling? become a distributor.

For the staff side of the same routine, the infection prevention training modules cover hand hygiene, PPE sequencing and between-patient cleaning in one place, and the clinic consumables hub groups the disposable lines that a single-room practice reorders most often.

Questions that go past a template — a label wording your importer uses, a record format a particular inspector asked for, or a retention period your regulator is challenging — belong with the compliance consultation desk rather than a generic contact form.

If you would rather test the routine than read about it, the biological indicator sample request is the simplest way to confirm that the incubator, the records and the staff habit all work together. The infection prevention study guide collects the rest of the guides for this setting.

Related Reading

Frequently Asked Questions

Are tympanometry tips single-use in Canada?

For the common commercial instruments, yes. Grason-Stadler states that TympStar ear tips are single-use only, and Interacoustics marks its AT235 with a no-reuse symbol and states that Sanibel ear tips are single use. Ontario's audiology guidelines add that items labelled single-use should not be reprocessed. A reusable tip is permissible only where the manufacturer supports it and the clinic reprocesses to the required level.

Is a disinfectant wipe enough for the probe between patients?

It is enough for the probe body, clip and cable, provided the wipe is a low-level disinfectant used for its stated contact time — one manufacturer explicitly instructs cleaning and disinfecting the probe dock between each patient. It is not enough for a reusable tip that has contacted the canal, which is semi-critical and needs high-level disinfection or sterilization.

Why do two Ontario documents classify the same tip differently?

The 2010 audiology infection control guidance names impedance probe tips as semi-critical; a 2020 pandemic-era practice advice table lists immittance tips as non-critical. Provincial infection control guidance resolves the conflict by requiring the higher level where the manufacturer and Spaulding disagree, so the semi-critical reading is the one to operate on.

Does cleaning the probe tubes daily replace the between-patient step?

No. Daily cleaning with a cleaning floss is specified to keep cerumen out of the measurement path, which is an accuracy requirement. The between-patient measure is a new tip plus a wipe of the hardware. Treating the floss as the infection control step leaves both obligations half met.

How should a clinic forecast tip consumption?

By appointments, not by rooms. One tip per tympanogram, so weekly consumption equals weekly tympanometry appointments plus a small allowance for retests and spoiled tips. Gloves are forecast on patient contacts rather than on tests, so the two lines move at different rates across a seasonal calendar.

CliniEco Medical holds Health Canada MDEL #35334. Audiology infection control in Canada is set provincially, and the provinces do not agree on the classification of a tympanometry tip. Where documents conflict, the stricter reading is the one to operate on. This page is general information about reprocessing equipment and consumables, not clinical or legal advice. Follow the instrument and sterilizer manufacturers' validated instructions; where this page and a maker's instructions differ, the maker's instructions govern. Health Canada MDEL #35334 covers our establishment licence; it is not a statement about any individual product.