Spore Testing for Autoclaves: What Dental Clinics Need to Know
It is Monday morning in a dental clinic in Mississauga. The office manager pulls a small plastic vial from the sterilization supply drawer, places it on the autoclave tray with a routine load of hand instruments, and runs a normal 20-minute cycle. When the cycle finishes, she labels the vial, puts it in the small incubator on the counter, and notes the date in the sterilization log. In 24 hours she will read the result: if the broth inside stays clear, the sterilizer passed. If it turns yellow, the clinic has a problem — and a documented one.
That routine is called a spore test, and it is the single most important check a clinic performs on its sterilizer. A spore test — also called a biological indicator (BI) test — is the only routine monitoring method that proves microorganisms were actually killed, rather than just confirming that the machine reached a temperature. Here is what Canadian dental clinics need to know about running spore tests correctly, reading the results, and keeping the records that inspectors ask for.
Biological indicators (24-hour, 25-pack)
Biological indicators (1-hour, 50-pack)
Looking for the supplies in this guide? We stock them in Canada and ship from Toronto.
Dental sterilization supplies → · Wholesale & clinic accounts · Free sterilization log tool · Ask a compliance specialist
What a Spore Test Actually Measures
Products referenced in this guide
- 24-Hour Biological Indicators — 25-Pack — $69.99
- 24-Hour Biological Indicators — 50-Pack — $109.99
- 24-Hour BI Incubator — Precision Dry Block — $379.00
A biological indicator contains a known population of highly resistant bacterial spores. For steam sterilizers, the spores are from Geobacillus stearothermophilus, a thermophilic organism that survives heat far better than the microorganisms typically found on used instruments. The standard that governs these organisms is ISO 11138-3, which specifies their resistance characteristics.
When the autoclave runs a cycle with the BI inside, one of two things happens. If the sterilizer delivered the correct combination of temperature, time, and steam quality, the spores are killed, and the BI's growth medium stays unchanged after incubation. If any critical parameter fell short, some spores survive, multiply during incubation, and produce a visible change in the medium — usually a color shift. The ISO 11138-1 series covers the manufacture, packaging, and performance testing of these indicators, which is why the BI products you buy should come from manufacturers that document compliance with it.
Because the spores are far more resistant than typical contamination, a passing spore test is strong evidence that the sterilizer is working correctly. A failing spore test is equally clear: the sterilization process failed, regardless of what the machine's display said.

How Often Biological Monitoring Is Required
The CDC Guidelines for Infection Control in Dental Health-Care Settings and the 2016 Summary of Infection Prevention Practices in Dental Settings both state that a biological indicator should be run at least once a week for every sterilizer. The CDC's Guideline for Disinfection and Sterilization in Healthcare Facilities goes further, recommending a BI in every load that contains implantable devices, and noting that the daily use of biological indicators is appropriate in some high-risk settings.
Canadian dental regulators align with this expectation. Provincial practice standards — including those from the Royal College of Dental Surgeons of Ontario — require documented sterilization monitoring — in Ontario, a biological indicator on each day a sterilizer is used; the weekly baseline is the CDC / AAMI ST79 standard in most US states. The CSA Group's Z314 series on device reprocessing reinforces the same message: mechanical monitoring is not enough; biological monitoring is part of a compliant quality assurance program.
In practical terms, the Ontario requirement means a biological indicator for each sterilizer on every day it is used, and for each type of cycle used that day. If your clinic has two autoclaves, each one needs its own test on each day it runs, and the results should be logged for each unit separately.
Step-by-Step: How to Run a Spore Test
Running a spore test correctly matters as much as running it at all. A BI placed in the wrong part of the load can give a false pass or a false fail. Follow this sequence:
- Check the expiry date and lot number. Never use an expired BI, and record the lot number so you can trace results back to the manufacturer if a failure occurs.
- Place the BI in the most challenging location. The standard advice is to put the BI inside a pack or at the location in the load that is hardest to sterilize — typically the densest or largest instrument pack, or the lower front of the chamber where steam penetration is most difficult.
- Run a normal full cycle. Do not run a special cycle for the test. The point is to verify the cycle you actually use for patient instruments.
- Remove the BI and label it. Write the date, sterilizer ID, cycle number, and load contents on the vial or a label.
- Incubate per the manufacturer's instructions. Most self-contained BIs are read at 24 hours; some are read at 48. Follow the exact temperature and time on the product insert — a BI read too early can produce a false negative.
- Read the result and record it. Compare the vial against the positive control that comes with the product. A negative control (unprocessed vial) should remain unchanged; a positive control should show growth, confirming the medium and incubator are working.
- Document everything. Log the date, sterilizer, cycle parameters, BI lot number, result, and the initials of the person who read it. This log is the record an inspector or public health officer will ask to see.
CliniEco Medical supplies a 24-hour self-contained biological indicator in a 25-pack format, which covers roughly five days of daily testing for one sterilizer and comes with a positive control for each pack.
How to Read the Results
After incubation, compare the test vial with the controls:
- Negative result (pass): The medium remains its original color. The spores were killed, and the sterilizer met the conditions needed to inactivate them.
- Positive result (fail): The medium changes color, indicating spore growth. The sterilization cycle did not kill the spores.
Rapid-readout BIs use a fluorescence or color-change system that can be read with the naked eye or a small reader. Always read at the time specified by the manufacturer. If the vial is left too long, even a negative test can eventually show growth because the medium is designed to support spore growth under the right conditions — a late reading can look like a failure, and an early reading can hide one.

What to Do If a Spore Test Is Positive
A positive spore test is not the time for guesswork. Treat it as a sterilization failure and follow a defined sequence:
- Stop using the sterilizer for patient instruments until the issue is resolved.
- Recall and reprocess the load. Instruments processed in the failed cycle are not sterile. Reprocess them through a functioning sterilizer, or quarantine them until the unit passes a repeat test.
- Troubleshoot the unit. Check the cycle parameters, door seal, steam supply, chamber cleanliness, and loading pattern. Many failures trace to a partially loaded chamber, a blocked drain, or steam quality issues rather than a mechanical breakdown.
- Run a repeat spore test. The CDC recommends two consecutive passing BIs before returning a sterilizer to service.
- Document the failure and the corrective action. A record that shows you identified the problem, reprocessed the load, and confirmed the fix is exactly what an inspector wants to see.
Common Mistakes That Invalidate a Spore Test
Clinics sometimes get a false confidence from a spore test that was never valid in the first place. Watch for these errors:
- Running a special cycle. If you run a longer or hotter cycle than usual, you are testing the special cycle, not your routine one.
- Placing the BI in the easiest location. A vial taped to the upper shelf may pass while the dense pack below fails.
- Reading too early. A self-contained BI read at 12 hours may not yet show growth that would appear at 24.
- Forgetting controls. Without a positive control, you cannot tell whether a negative result means the spores were killed or the medium never supported growth.
- Running the test in an empty chamber. A lightly loaded chamber sterilizes more easily than a full, real load.
- Not logging results. A passing test that was never recorded does not exist for inspection purposes.
Documentation: The Log That Saves You
Spore test results are part of the sterilization record that provincial inspectors and public health officers review. A complete entry includes the date, sterilizer identification, cycle type, load contents, BI lot number, incubation time, result, and operator initials. Keeping this log in a consistent format — whether paper or digital — makes the monitoring routine faster and makes inspections straightforward.
Pair the biological indicator with the rest of the monitoring chain: Class 4 sterilization pouches with internal chemical indicators for every pack, and autoclave indicator tape on external wraps. The full set — chemical, biological, and mechanical monitoring — is what the CDC's disinfection and sterilization guideline calls a complete monitoring program.
Choosing a Biological Indicator Product
When you select a BI product for your clinic, check four things:
- Rapid readout vs. traditional. A 24-hour self-contained system lets you release instruments the next morning instead of waiting 48 hours.
- Incubator requirements. Some products need a specific dry-block incubator; confirm the product matches the equipment you have or buy the pair together.
- Controls included. The pack should include positive controls so every test batch is verified.
- Documentation from the manufacturer. Look for a supplier who can provide technical data and who works with licensed manufacturers. CliniEco Medical is a licensed medical device establishment (MDEL #35334) and its sterilization monitoring line is supplied from licensed manufacturing sources.
For clinics building a complete program, CliniEco Medical's sterilization monitoring collection brings pouches, rolls, tape, and biological indicators together under one supplier — which also simplifies the paperwork when an inspector asks where your monitoring products came from.
What each pack costs per test
| Pack | Price | Cost per test | Order |
|---|---|---|---|
| 24-hour self-contained BI, 25-pack | $69.99 | $2.80 | 24-Hour BI, 25-Pack |
| 24-hour self-contained BI, 50-pack | $109.99 | $2.20 | 24-Hour BI, 50-Pack |
| 24-hour self-contained BI, 100-pack | $189.00 | $1.89 | 24-Hour BI, 100-Pack |
| 3-hour fluorescence BI, 50-pack | $249.99 | $5.00 | 3-Hour Rapid BI, 50-Pack |
Prices are the list prices we sell at, not estimates: a 24-hour vial costs $2.80 in the 25-pack, $2.20 in the 50-pack and $1.89 in the 100-pack, so the 100-pack is the lowest cost per test we stock.
Which sterilizers and incubators these products fit
| Product | Sterilizer cycles it is validated for | Reader or incubator you need |
|---|---|---|
| 24-hour self-contained BI (25, 50 or 100-pack) | 121 °C gravity-displacement and 132–135 °C dynamic-air-removal (pre-vacuum) steam cycles — any make of steam autoclave, benchtop or floor model, as long as the cycle type matches the product insert | A 55–60 °C dry-block incubator; 24-well and 40-well units both take standard self-contained vials |
| 3-hour fluorescence BI (50-pack) | The same steam cycles, read by enzyme fluorescence | A compatible 4-well fluorescence reader, which returns the result in 3 hours |
| Class 5 integrator, indicator tape, pouches | Process monitoring inside every load — nothing is incubated | None |
The fit question is settled by the cycle type on the product insert, not by the sterilizer brand: a biological indicator validated for 121 °C gravity and 132–135 °C vacuum cycles can be run in any steam sterilizer running those cycles, and the vial format — not the sterilizer — decides which incubator or reader you need.
How many packs to buy at once. Tests in a working week = sterilizers × days each sterilizer is used × cycle types used that day. In Ontario that is one biological indicator per sterilizer on every day it is used, plus one per cycle type that day (RCDSO); outside Ontario the CDC / AAMI ST79 baseline is one test for each sterilizer every seven days. A single-autoclave Ontario practice open five days needs 5 tests a week, about 22 a month, so one 25-pack covers five weeks; two autoclaves at six days and two cycle types need 24 a week, about 104 a month, so one 100-pack covers four weeks. Add one positive control per incubation batch.
Stock and shipping. The 24-hour and 3-hour packs are held in stock and ship from our Toronto warehouse. We do not quote a delivery date — carrier transit time is outside our control.
Free Printable: Spore Test Log
Print this page (Ctrl+P, or ⌘P on a Mac) or save it as a PDF. Post it where the work happens.
One line per test. Keep the log with the sterilizer.| Date | Sterilizer ID | BI lot # | Result | Action if fail | |
|---|---|---|---|---|---|
| Date | Sterilizer ID | BI lot # | Result (pass/fail) | Action if fail | |
FAQ: Spore Testing for Autoclaves
How often should a dental clinic run a spore test?
In Ontario, one biological indicator for each sterilizer on every day it is used, and one for each type of cycle used that day — the RCDSO requirement. United States guidance (CDC, AAMI ST79) sets a weekly baseline instead. If you have multiple autoclaves, test each one separately.
What happens if a spore test comes back positive?
Stop using the sterilizer, recall and reprocess the load, troubleshoot the unit, and run two consecutive passing spore tests before returning it to service. Document the failure and corrective actions.
Do I need a positive control with every spore test?
Yes. A positive control confirms that the growth medium and incubator are working. Without it, a negative result could mean the medium failed rather than the sterilizer succeeded.
Can I read a rapid-readout biological indicator early?
No. Read at the time specified by the manufacturer — usually 24 hours. Reading early can produce a false negative; reading late can show growth in a vial that was actually negative.
Where should the biological indicator be placed in the load?
In the most difficult-to-sterilize location, typically inside the densest pack or at the lower front of the chamber where steam penetration is hardest. Testing the easiest spot tests nothing.
What organism is used in autoclave spore tests?
Geobacillus stearothermophilus, a heat-resistant bacterium specified in ISO 11138-3 for moist heat sterilization. It is far more resistant than typical clinical contamination.
Is a spore test the same as a chemical indicator?
No. A chemical indicator confirms that a condition (like temperature) was present; a spore test proves microorganisms were killed. They are complementary, and a compliant program uses both.
Do veterinary and medical clinics need spore testing too?
Yes. Any facility that reprocesses instruments in a steam sterilizer — dental, medical, veterinary, or aesthetic — should monitor every sterilizer on each day it is used wherever the local regulator requires daily monitoring, and never less than the weekly baseline set out in United States guidance.
Browse our autoclave and ultrasonic equipment.
Related reading: browse the sterilization compliance hub for more spore-testing resources.Related: Sterilization subscription: daily BI + integrators + pouches + tape, from $69/month.
read the foundational SAL explainer before choosing your monitoring products.
References
- ISO 11138-1 — Sterilization of health care products: Biological indicators. https://www.iso.org/standard/66448.html
- ISO 11138-3 — Biological indicators for moist heat sterilization (Geobacillus stearothermophilus). https://www.iso.org/standard/67353.html
- CDC. Guidelines for Infection Control in Dental Health-Care Settings, 2003 (MMWR RR-17). https://www.cdc.gov/mmwr/preview/mmwrhtml/rr5217a1.htm
- CDC. Summary of Infection Prevention Practices in Dental Settings, 2016. https://www.cdc.gov/oralhealth/infectioncontrol/index.html
- CDC. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008. https://www.cdc.gov/infectioncontrol/guidelines/disinfection/
- AAMI ST79 — Comprehensive guide to steam sterilization and sterility assurance. https://www.aami.org/
- Health Canada. Medical devices (regulation overview). https://www.canada.ca/en/health-canada/services/drugs-health-products/medical-devices.html
- Health Canada. Medical device establishment licences (GUI-0064). https://www.canada.ca/en/health-canada/services/drugs-health-products/medical-devices/licences.html
- CSA Group. Z314 series — Reprocessing of medical devices in Canada. https://www.csagroup.org/
- PHAC. Routine Practices and Additional Precautions. https://www.canada.ca/en/public-health/services/infection-prevention-control-policies/routine-practices-additional-precautions.html
- Royal College of Dental Surgeons of Ontario (RCDSO) — Infection prevention and control standards. https://www.rcdso.org/
- Canadian Dental Association (CDA) — Position statements and infection control resources. https://www.cda-adc.ca/
- WHO. Bulletin of the World Health Organization — Health care-associated infection publications. https://www.who.int/publications/journals/bulletin
- Journal of Hospital Infection — Sterilization and disinfection research. https://www.sciencedirect.com/journal/journal-of-hospital-infection
- American Journal of Infection Control — Infection prevention evidence. https://www.sciencedirect.com/journal/american-journal-of-infection-control
- Emerging Infectious Diseases (CDC) — Pathogen transmission research. https://wwwnc.cdc.gov/eid/
- sterilization compliance log sheet (free tool)
CliniEco Medical is a licensed medical device establishment (MDEL #35334). This article is for informational purposes; always follow your sterilizer manufacturer's instructions, the BI product insert, and your provincial regulator's requirements.
Stock Your Spore Test Program
CliniEco Medical supplies 24-hour self-contained biological indicators, Class 4 sterilization pouches, autoclave indicator tape, and sterilization rolls to Canadian dental, veterinary, and medical clinics. For bulk pricing or a facility quote, contact us at clinieco.ca/pages/contact — a real person responds within one business day.
Related reading: indicators, incubator and reader — priced separately — a 24-well dry-block incubator is $379, nothing is bundled, and nothing is locked to us.
Monitoring your sterilizer? Pick the BI format that matches your workflow.
Ontario dental offices must run a biological indicator every day the sterilizer is used (RCDSO). Choose the readout speed that fits your schedule:
- 24-hour self-contained BIs — 25, 50 and 100-count packs for daily monitoring
- 1-, 3- and 20-minute rapid-readout BIs — same-day release when you cannot wait overnight
- 4-well fluorescence reader — incubates and reads rapid-readout BIs in one bench instrument
Shop 24-hour biological indicators See the 4-well rapid reader
CliniEco Medical · Health Canada MDEL #35334 · Request a wholesale quote
Spore testing is one thread of a larger monitoring record. A free sterilization self-check runs through the same records and checks an inspector is most likely to look for, by sector — roughly two minutes, printable, no sign-up.
Setting up or reviewing a spore test program? Send us your sterilizer and cycle types and we will walk through the monitoring schedule your records need. Ask a sterilization compliance specialist.
Related reading
- Do You Need a Biological Indicator for Every Load? Ontario Answers
- Rapid Readout Biological Indicator Readers: Every Major Brand Compared (2026 Canada Price Guide)
- BI vs Chemical Indicators: What's the Difference in Sterilization Monitoring?
- The Real Annual Cost of Biological Monitoring: Locked Ecosystems vs Open-System BIs in Canadian Dental Clinics
- Dental practice supplies and sterilization monitoring — what Ontario dental offices reorder every month, and the monitoring records behind it.
Where to Buy Sterilization Indicators and Spore Tests in Canada
CliniEco Medical supplies chemical indicators, integrating indicators and self-contained biological indicators to clinics, laboratories and care homes across Canada, shipped from Ontario.
| Type | What it confirms | Order |
|---|---|---|
| Class 5 integrating indicator | All critical parameters were met — used inside every pack | Class 5 Integrators, 100-Pack |
| Biological indicator (spore test) | The actual kill — the only test that proves sterilization | 24-Hour, 25-Pack |
| Bowie-Dick test pack | Air removal in pre-vacuum sterilizers only | Bowie-Dick, 6-Pack |
| Indicator tape | External process indicator — the load was processed | Autoclave Tape, 3-Roll Set |
| Not sure which applies? | Run one week in your own workflow first | 24-hour biological indicator 25-pack |
Round Out Your Sterilization Setup
A biological indicator is one link in the chain — these are the pieces it works with:
- Sterilization roll — 50 m pre-cut sheets for larger instruments
- Constant-heat pouch sealer — 200 °C, 12 mm seal — for consistent pouch closures
Ontario and GTA orders ship free over $99; elsewhere in Canada over $199 ($15 flat below). Bulk and standing orders go through a wholesale account. Print a blank sterilization log sheet from our free generator, or book a sterilization compliance review.
0 comments